Clinical UM Guideline |
Subject: Dynamic Low-Load Prolonged-Duration Stretch Devices | |
Guideline #: CG-DME-39 | Publish Date: 06/28/2024 |
Status: Reviewed | Last Review Date: 05/09/2024 |
Description |
This document addresses dynamic low-load prolonged-duration stretch (LLPS) devices for the treatment of joint stiffness and contracture.
Dynamic LLPS (dynamic splinting) devices are designed to provide a low load, prolonged stretch to joints that have reduced range of motion secondary to immobilization, surgery, contracture, fracture, dislocation, or a number of additional non-traumatic disorders. The objective of stretch therapy is to improve range of motion without compromising the stability and quality of the connective tissue and joint.
For additional information, please see the following documents:
Clinical Indications |
Medically Necessary:
Dynamic LLPS devices are considered medically necessary when criteria A., B and C below have been met:
Dynamic LLPS devices for the postoperative management of tendon repair, referred to as "protected motion", to limit motion and protect the underlying surgical repair are considered medically necessary.
Replacement interface material used with the dynamic LLPS device is considered medically necessary when both the following criteria have been met:
Not Medically Necessary:
If there is no significant improvement after 4 months of use, dynamic LLPS devices are considered not medically necessary under any circumstance, including but not limited to for individuals unable to benefit from standard physical or occupational therapy modalities because of an inability to exercise.
Dynamic LLPS devices are considered not medically necessary for use on any other joint or for any other condition not listed above, including but not limited to the management of chronic joint stiffness or chronic or fixed contractures.
Coding |
The following codes for treatments and procedures applicable to this guideline are included below for informational purposes. Inclusion or exclusion of a procedure, diagnosis or device code(s) does not constitute or imply member coverage or provider reimbursement policy. Please refer to the member's contract benefits in effect at the time of service to determine coverage or non-coverage of these services as it applies to an individual member.
When services may be Medically Necessary when criteria are met:
HCPCS |
|
E1399 | Durable medical equipment, miscellaneous [when specified as a dynamic LLPS device, for example Carpal Tunnel Dynasplint System] |
E1800 | Dynamic adjustable elbow extension/flexion device, includes soft interface material |
E1802 | Dynamic adjustable forearm pronation/supination device, includes soft interface material |
E1805 | Dynamic adjustable wrist extension/flexion device, includes soft interface material |
E1810 | Dynamic adjustable knee extension/flexion device, includes soft interface material |
E1812 | Dynamic knee, extension/flexion device with active resistance control |
E1820 | Replacement soft interface material, dynamic adjustable extension/flexion device |
E1825 | Dynamic adjustable finger extension/flexion device, includes soft interface material |
|
|
ICD-10 Diagnosis |
|
| All diagnoses |
When services are Not Medically Necessary:
For the procedure codes listed above when criteria are not met or for situations designated in the Clinical Indications section as not medically necessary.
When services are also Not Medically Necessary:
For the following procedure codes, or when the code describes a procedure or situation designated in the Clinical Indications section as not medically necessary.
HCPCS |
|
E1815 | Dynamic adjustable ankle extension/flexion device, includes soft interface material |
E1830 | Dynamic adjustable toe extension/flexion device, includes soft interface material |
E1840 | Dynamic adjustable shoulder flexion/abduction/rotation device, includes soft interface material |
|
|
ICD-10 Diagnosis |
|
| All diagnoses |
Discussion/General Information |
Dynamic LLPS devices are spring-loaded, adjustable systems which provide a low-load prolonged stretch. Dynamic LLPS devices provide a sustained, set level of tension using integrated springs. These devices are set at a fixed joint angle and worn over long periods (6-8 hours/day or overnight). Dynamic LLPS units provide extension as well as flexion and are available for various joints including the elbows, wrists, fingers, knees, ankles, and toes. These devices are being marketed for the treatment of joint stiffness due to immobilization or limited range of motion (ROM) as a result of various conditions, including, but not limited to, fractures, dislocations, tendon and ligament repairs, joint arthroplasties, total knee replacements, burns, rheumatoid arthritis, hemophilia, tendon releases, head trauma, spinal cord injuries, cerebral palsy, multiple sclerosis, and other traumatic and non-traumatic disorders. The purpose of the dynamic LLPS device is to restore ROM to a joint without diminishing the stability and quality of the connective tissue and joint. Dynamic LLPS devices are frequently used during the post-operative period to prevent or treat motion stiffness or decreased ROM in the knee, elbow, wrist or finger. These devices are generally not used in other joints such as the hip, ankle or foot. Examples of this type of device include but are not limited to:
Rives and colleagues (1992) reported the results of a prospective, non-randomized study in which a total of 23 proximal interphalangeal joints that were severely contracted (approximately 45 degrees) as a result of Dupuytren's disease underwent operative correction and 6 months of dynamic extension splinting. Proximal interphalangeal joint extension was measured preoperatively and postoperatively at 3-month intervals for 1 year and at 6-month intervals thereafter. Mean follow-up was 2 years (minimum, 1 year). Overall, at 2 years, 44% improvement in proximal interphalangeal joint extension was noted. Mean improvement of 59% in proximal interphalangeal joint extension was noted in the participants who complied with the postoperative dynamic extension splinting program. Participants who were noncompliant demonstrated a 25% improvement in proximal interphalangeal joint extension. The difference in values between participants who were compliant and those who were not was statistically significant. Other factors such as the severity of contracture, the digit involved, and the necessity for capsular release were not significantly related to outcome. The authors concluded that soft tissue responds to continuous dynamic extension stresses and can be remodeled over time.
In a descriptive study, Nuismer and colleagues (1997) retrospectively evaluated a convenience sample of participants at multiple sites using LLPS orthoses for contracture management. The records of 17 participants from skilled nursing facilities, hand clinics, and hospitals were reviewed. There were a total of 18 contractures (2 wrists, 12 elbows, 4 knees) secondary to neurological and orthopedic causes. Chart review focused on participant demographic information, ROM, functional outcomes, and wear schedules. The authors found that the use of LLPS orthoses significantly increased ROM for the whole sample, mean 28.61 degrees (range, -10-66, p<0.001), which in turn significantly improved the participants' functional outcomes. When the sample was divided into two pathology groups to compare a predominately geriatric population with neurological pathologies to a somewhat younger population with a history of musculoskeletal pathology, both groups showed a significant gain in ROM with the use of the LLPS orthoses. The authors concluded that the use of LLPS orthoses for contracture management can mediate the losses in ROM and function that occur with joint contractures.
Lindenhovius and colleagues (2012) compared the use of dynamic and static progressive (turnbuckle) splints to help stretch a contracted elbow capsule and regain motion after elbow trauma. In a prospective randomized controlled trial the authors tested the null hypothesis that there is no difference in improvement of motion and Disabilities of the Arm, Shoulder and Hand (DASH) scores between static progressive and dynamic splinting. A total of 66 participants with posttraumatic elbow stiffness were enrolled in a prospective randomized trial: 35 in the static progressive and 31 in the dynamic cohort. Elbow function was measured at enrollment and at 3, 6, and 12 months later and the DASH questionnaire at enrollment and at the 6- and 12-month evaluation. A total of 3 participants asked to be switched to static progressive splinting. The analysis was done according to intention-to-treat principles and with use of mean imputation for missing data. There were no significant differences in flexion arc at any time point. Improvement in the arc of flexion (dynamic versus static) averaged 29 degrees versus 28 degrees at 3 months (p=0.87), 40 degrees versus 39 degrees at 6 months (p=0.72), and 47 degrees versus 49 degrees at 12 months after splinting was initiated (p=0.71). The average DASH score (dynamic versus static) was 50 versus 45 points at enrollment (p=0.52), 32 versus 25 points at 6 months (p<0.05), and 28 versus 26 points at 12 months after enrollment (p=0.61). The authors concluded that posttraumatic elbow stiffness can improve with exercises and dynamic or static splinting over a period of 6 to 12 months, and patience is warranted. There were no significant differences in improvement in motion between static progressive and dynamic splinting protocols.
Although there is limited data from controlled trials in the published peer reviewed literature, this technology is widely used by orthopedists as well as occupational and physical therapists for selected populations. Additional input from medical practitioners practicing in relevant clinical areas support the role of dynamic splinting in the clinical settings outlined under the Clinical Indications section of this document.
Definitions |
Protected motion: A clinical concept which recognizes the importance of protecting damaged tissues as well as the benefits of controlled motion during the early stages of connective tissue repair.
References |
Peer Reviewed Publications:
Index |
Advance Dynamic ROM
Dynamic Splinting
Dynasplint
JAS Dynamic
Pro-Glide
Spring Loaded Dynamic Splinting
The use of specific product names is illustrative only. It is not intended to be a recommendation of one product over another, and is not intended to represent a complete listing of all products available.
History |
Status | Date | Action |
Reviewed | 05/09/2024 | Medical Policy & Technology Assessment Committee (MPTAC) review. Updated review date, Discussion/General Information and History sections. |
Revised | 05/11/2023 | MPTAC review. Reformatted Clinical Indications, updated review date and History sections. |
Reviewed | 05/12/2022 | MPTAC review. Updated review date and History section. |
Reviewed | 05/13/2021 | MPTAC review. Updated Discussion/General Information and History sections. Reformatted Coding section. |
Reviewed | 05/14/2020 | MPTAC review. Updated History section. |
Revised | 06/06/2019 | MPTAC review. Updated the Discussion/General Information and Index sections. References to the “Elite Seat” device moved to DME.00038 Static Progressive Stretch (SPS) and Patient-Actuated Serial Stretch (PASS) devices to align with the proper classification of the device. In the Not Medically Necessary Position Statement, changed “four” to “4”. Updated Coding section to add E1399 NOC code. |
Reviewed | 03/21/2019 | MPTAC review. Updated the Discussion/General Information and Index sections. |
Reviewed | 03/22/2018 | MPTAC review. The document header wording updated from “Current Effective Date” to “Publish Date.” |
Reviewed | 05/04/2017 | MPTAC review. Updated formatting in the Clinical Indications section. |
New | 05/05/2016 | MPTAC review. Initial document development. |
Federal and State law, as well as contract language, and Medical Policy take precedence over Clinical UM Guidelines. We reserve the right to review and update Clinical UM Guidelines periodically. Clinical guidelines approved by the Medical Policy & Technology Assessment Committee are available for general adoption by plans or lines of business for consistent review of the medical necessity of services related to the clinical guideline when the plan performs utilization review for the subject. Due to variances in utilization patterns, each plan may choose whether to adopt a particular Clinical UM Guideline. To determine if review is required for this Clinical UM Guideline, please contact the customer service number on the member's card.
Alternatively, commercial or FEP plans or lines of business which determine there is not a need to adopt the guideline to review services generally across all providers delivering services to Plan’s or line of business’s members may instead use the clinical guideline for provider education and/or to review the medical necessity of services for any provider who has been notified that his/her/its claims will be reviewed for medical necessity due to billing practices or claims that are not consistent with other providers, in terms of frequency or in some other manner.